TECHNICAL SPECIFICATION
· SYSTEM FUNCTION: Automatic Selective Access Flow System
· THROUGHPUT: 200 tests/hour
· METHODOLOGY: End point\Kinetic\Initial Rate\Bichromatic-Differential
· PROGRAMMING: User defined profiles and calculation chemistries |
|
SAMPLE HANDLING
· SAMPLE TRAY: Four racks for a total of 40 positions
· SAMPLE VOLUME: From 3 ul to 100 ul with 1 ul increment
· ACCURACY: 99%
· MATERIALS: Teflon&Glass
|
|
REAGENT HANDLING
· REAGENT TRAY: 18 positions in refrigerated reagent compartment(4-10℃)
· REAGENT VOLUME: 5-900 ul with 1 ul increment
|
|
ANALYZER SYSTEM
· CUVETTE BODY: Stainless steel & Quartz
· CUVETTE VOLUME: 50 ul
· OPERATING TEMPERATURE: 37℃
· ACCURACY: ± 0.1℃
|
|
|
|
| |
|
| |
|
|
|
OPTICAL SYSTEM
· LIGHT SOURCE: Halogen-tungsten lamp(6V/10W)
· PHOTOMETER: Seven narrow bandwidth interference filters
· WAVELENGTH: 340-800nm
· ABSORPTION RANGE: 0.000-3.000A
· RESOLUTION: ±0.0001A
|
|
|
|
|
|
WORKING CONDITIONS
· POWER REQUIREMENT: 230 V - 50 Hz / 115 V - 60 Hz 400 Watt
· HUMIDITY: 35-80%, no condensation
· DIMENSIONS: 80 cm x 70 cm x 40 cm
· WEIGHT: 50Kg(Not including the cabinet)
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
 |
· Exports South Korea, Vietnam |
|
|
 |
|
|
|
|
|
|